AIIMS-led team develops sensitive tests for pulmonary, pleural TB

Sensitivity for pulmonary TB and pleural TB were 94% and 93%, respectively.

November 10, 2018 06:18 pm | Updated 06:20 pm IST

Careful test:  The sputum should first be made into a uniform suspension before applying the test. (Photo used for illustrative purposes only.)

Careful test: The sputum should first be made into a uniform suspension before applying the test. (Photo used for illustrative purposes only.)

A highly sensitive diagnostic test for pulmonary TB and pleural TB has been developed by a multi-institutional team led by Jaya Sivaswami Tyagi from the Department of Biotechnology at AIIMS. The diagnostic test makes use of a DNA aptamer (a small single-stranded DNA molecule that binds to a specific target molecule) that shows high binding affinity to a TB antigen. Sputum samples were used for diagnosing pulmonary TB while pleural fluid was used for diagnosing pleural TB.

The sensitivity of the diagnostic test for pulmonary TB and pleural TB was 94% and 93%, respectively. Specificity was 100% for pulmonary TB and 98% for pleural TB. The results were published in the journals ACS Infectious Diseases and Analytical Biochemistry.

The laboratory-based aptamer test for diagnosing pulmonary TB takes about five hours to turn in the results while the portable, rapid test takes just 30 minutes. “The lab-based test has several steps and so takes time while the rapid test is an electrochemical sensor-based detection which is a direct detection and so takes less time,” says Surabhi Lavania from AIIMS and first author of the pulmonary TB paper and coauthor of the other paper. At 92.3% and 91.2%, the sensitivity and specificity of the portable test is a bit inferior compared with the lab-based test.

In the case of pleural TB, the aptamer-based lab diagnosis was far superior to even Xpert. While the sensitivity was about 22% with Xpert, it was about 93% with aptamer. “Xpert relies on the presence of bacteria, which is nearly absent in pleural fluid, while the aptamer-based testing relies on bacterial antigen and so has high sensitivity,” says Pooja Kumari from AIIMS and first author of the paper on pleural TB test.

The sensitivity of the lab-based and portable aptamer-based test for pulmonary TB is far superior compared with smear microscopy, X-ray and even ELISA.

Cost-effective

“The portable test is cheaper and can be used for screening pulmonary TB and active case finding in high-risk groups,” says Prof. Tyagi.

The HspX antigen present in sputum samples of people with pulmonary TB is made use of for diagnosis. The team had used the same antigen present in the cerebrospinal fluid samples for diagnosing TB meningitis (published in September 2018 in Tuberculosis ). The researchers made use of a particular aptamer (H63SL2-M6) that detects the antigen in sputum samples for pulmonary TB diagnosis.

“Sputum is a difficult sample to handle due to the presence of mucus. The sputum should first be made into a uniform suspension before applying the test. So [we] must prepare the sample and this takes time,” says Prof. Tyagi.

In the case of the lab-based diagnosis, the sputum sample is immobilised on a plate and the aptamer is added. If the sample is positive for pulmonary TB then it will contain the HspX antigen and the aptamer binds to it. An enzyme (horseradish peroxidise) that binds to the aptamer (through biotin-streptavidin affinity) is then added followed by a substrate. A change in substrate colour is seen indicating that the sample is positive for pulmonary TB. If the sample is negative for pulmonary TB, then even the first step of the aptamer binding to the sample does not take place and the colour change does not occur.

Relying on HspX antigen to diagnose pleural TB in a lab setting was found be effective and highly sensitive (93%).

Rapid diagnosis

To make the pulmonary TB diagnosis portable, the researchers bound the aptamer to an electrode coated with gold nanoparticles. “When the sample containing the HspX antigen is added, the aptamer binds to the antigen and undergoes a structural change, which is read out in the electrical signal,” says Dr. Tarun Kumar Sharma from Translational Health Science and Technology Institute (THSTI), Faridabad, and the other corresponding author of the papers. “There is a drop in electrical signal if the sample is positive and no change in signal when the sample is negative.”

The electrode is already immobilised with the aptamer, so only the liquefied sputum has to be added and the reaction is instantaneous. “Most time is taken to liquefy the sputum sample than the testing per se,” says Prof. Tyagi. No sample processing is needed in the case of pleural TB samples.

“We are also trying to adapt this aptamer on an electrode for rapid diagnosis of pleural TB,” says Dr. Sharma.

“We have applied for a patent for the aptamer reagent and have licensed it to AptaBharat Innovation Pvt. Ltd, a start-up at THSTI. We soon plan to validate the tests on well characterised panel of specimens and field evaluation,” says Prof. Tyagi.

0 / 0
Sign in to unlock member-only benefits!
  • Access 10 free stories every month
  • Save stories to read later
  • Access to comment on every story
  • Sign-up/manage your newsletter subscriptions with a single click
  • Get notified by email for early access to discounts & offers on our products
Sign in

Comments

Comments have to be in English, and in full sentences. They cannot be abusive or personal. Please abide by our community guidelines for posting your comments.

We have migrated to a new commenting platform. If you are already a registered user of The Hindu and logged in, you may continue to engage with our articles. If you do not have an account please register and login to post comments. Users can access their older comments by logging into their accounts on Vuukle.